Programmable multi-pattern wide frequency range electrical stimulator

Abbas Orand, Genichi Tanino, Hiroyuki Miyasaka, K. Takeda, S. Sonoda

研究成果: ジャーナルへの寄稿学術論文査読

抄録

In this paper, a programmable, multi-pattern, wide frequency and duty cycle range electrical stimulator is presented. Using a programmable microcontroller, two waves of carrier and modulating sources are produced. By modulating the two sources, 3 rectangular, triangular and sinusoidal stimulating patterns are produced. The frequency range of the carrier is fixed at 2.5 kHz and the carrier source frequency can be adjusted between 1 and 500 Hz. The duty cycle of both sources can be adjusted between 10% and 90%.

本文言語英語
ジャーナルJournal of Electrical Engineering
16
3
出版ステータス出版済み - 2016
外部発表はい

All Science Journal Classification (ASJC) codes

  • 電子工学および電気工学