T1 - Quantifying the performance of two different types of commercial software programs for 3D patient dose reconstruction for prostate cancer patients
T2 - Machine log files vs. machine log files with EPID images
AU - Kadoya, Noriyuki
AU - Kon, Yoshio
AU - Takayama, Yoshiki
AU - Matsumoto, Takuya
AU - Hayashi, Naoki
AU - Katsuta, Yoshiyuki
AU - Ito, Kengo
AU - Chiba, Takahito
AU - Dobashi, Suguru
AU - Takeda, Ken
AU - Jingu, Keiichi
N1 - Publisher Copyright:
© 2017 Associazione Italiana di Fisica Medica
PY - 2018/1
Y1 - 2018/1
N2 - We clarified the reconstructed 3D dose difference between two different commercial software programs (Mobius3D v2.0 and PerFRACTION v1.6.4). Five prostate cancer patients treated with IMRT (74 Gy/37 Fr) were studied. Log files and cine EPID images were acquired for each fraction. 3D patient dose was reconstructed using log files (Mobius3D) or log files with EPID imaging (PerFRACTION). The treatment planning dose was re-calculated on homogeneous and heterogeneous phantoms, and log files and cine EPID images were acquired. Measured doses were compared with the reconstructed point doses in the phantom. Next, we compared dosimetric metrics (mean dose for PTV, rectum, and bladder) calculated by Mobius3D and PerFRACTION for all fractions from five patients. Dose difference at isocenter between measurement and reconstructed dose for two software programs was within 3.0% in both homogeneous and heterogeneous phantoms. Moreover, the dose difference was larger using skip arc plan than that using full arc plan, especially for PerFRACTION (e.g., dose difference at isocenter for PerFRACTION: 0.34% for full arc plan vs. −4.50% for skip arc plan in patient 1). For patients, differences in dosimetric parameters were within 1% for almost all fractions. PerFRACTION had wider range of dose difference between first fraction and the other fractions than Mobius3D (e.g., maximum difference: 0.50% for Mobius3D vs. 1.85% for PerFRACTION), possibly because EPID may detect some types of MLC positioning errors such as miscalibration errors or mechanical backlash which cannot be detected by log files, or that EPID data might include image acquisition failure and image noise.
AB - We clarified the reconstructed 3D dose difference between two different commercial software programs (Mobius3D v2.0 and PerFRACTION v1.6.4). Five prostate cancer patients treated with IMRT (74 Gy/37 Fr) were studied. Log files and cine EPID images were acquired for each fraction. 3D patient dose was reconstructed using log files (Mobius3D) or log files with EPID imaging (PerFRACTION). The treatment planning dose was re-calculated on homogeneous and heterogeneous phantoms, and log files and cine EPID images were acquired. Measured doses were compared with the reconstructed point doses in the phantom. Next, we compared dosimetric metrics (mean dose for PTV, rectum, and bladder) calculated by Mobius3D and PerFRACTION for all fractions from five patients. Dose difference at isocenter between measurement and reconstructed dose for two software programs was within 3.0% in both homogeneous and heterogeneous phantoms. Moreover, the dose difference was larger using skip arc plan than that using full arc plan, especially for PerFRACTION (e.g., dose difference at isocenter for PerFRACTION: 0.34% for full arc plan vs. −4.50% for skip arc plan in patient 1). For patients, differences in dosimetric parameters were within 1% for almost all fractions. PerFRACTION had wider range of dose difference between first fraction and the other fractions than Mobius3D (e.g., maximum difference: 0.50% for Mobius3D vs. 1.85% for PerFRACTION), possibly because EPID may detect some types of MLC positioning errors such as miscalibration errors or mechanical backlash which cannot be detected by log files, or that EPID data might include image acquisition failure and image noise.
UR - http://www.scopus.com/inward/record.url?scp=85042077418&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85042077418&partnerID=8YFLogxK
U2 - 10.1016/j.ejmp.2017.12.018
DO - 10.1016/j.ejmp.2017.12.018
M3 - Article
C2 - 29472083
AN - SCOPUS:85042077418
SN - 1120-1797
VL - 45
SP - 170
EP - 176
JO - Physica Medica
JF - Physica Medica
ER -