T1 - Quantitatively assessed CT imaging measures of pulmonary interstitial pneumonia
T2 - Effects of reconstruction algorithms on histogram parameters
AU - Koyama, Hisanobu
AU - Ohno, Yoshiharu
AU - Yamazaki, Youichi
AU - Nogami, Munenobu
AU - Kusaka, Akiko
AU - Murase, Kenya
AU - Sugimura, Kazuro
N1 - Funding Information:
We thank Masatoshi Kondo (CT Clinic. Scientist, Philips Electronics Japan), Keisuke Fukuda (Technical & Support Department of Philips Electronics Japan), Yumiko Onishi, M.D., Minoru Konishi, R.T., Masaru Yoshii, R.T, Masahiko Fujii, M.D. (Division of Radiology, Kobe University Hospital), Keiko Matsumoto M.D. and Daisuke Takenaka, M.D. (Department of Radiology, Kobe University Graduate School of Medicine) for their contribution to this work. This work was supported by Philips Medical Systems.
Copyright:
Copyright 2010 Elsevier B.V., All rights reserved.
PY - 2010/4
Y1 - 2010/4
N2 - This study aimed the influences of reconstruction algorithm for quantitative assessments in interstitial pneumonia patients. A total of 25 collagen vascular disease patients (nine male patients and 16 female patients; mean age, 57.2 years; age range 32-77 years) underwent thin-section MDCT examinations, and MDCT data were reconstructed with three kinds of reconstruction algorithm (two high-frequencies [A and B] and one standard [C]). In reconstruction algorithm B, the effect of low- and middle-frequency space was suppressed compared with reconstruction algorithm A. As quantitative CT parameters, kurtosis, skewness, and mean lung density (MLD) were acquired from a frequency histogram of the whole lung parenchyma in each reconstruction algorithm. To determine the difference of quantitative CT parameters affected by reconstruction algorithms, these parameters were compared statistically. To determine the relationships with the disease severity, these parameters were correlated with PFTs. In the results, all the histogram parameters values had significant differences each other (p < 0.0001) and those of reconstruction algorithm C were the highest. All MLDs had fair or moderate correlation with all parameters of PFT (-0.64 < r < -0.45, p < 0.05). Though kurtosis and skewness in high-frequency reconstruction algorithm A had significant correlations with all parameters of PFT (-0.61 < r < -0.45, p < 0.05), there were significant correlations only with diffusing capacity of carbon monoxide (DLco) and total lung capacity (TLC) in reconstruction algorithm C and with forced expiratory volume in 1 s (FEV1), DLco and TLC in reconstruction algorithm B. In conclusion, reconstruction algorithm has influence to quantitative assessments on chest thin-section MDCT examination in interstitial pneumonia patients.
AB - This study aimed the influences of reconstruction algorithm for quantitative assessments in interstitial pneumonia patients. A total of 25 collagen vascular disease patients (nine male patients and 16 female patients; mean age, 57.2 years; age range 32-77 years) underwent thin-section MDCT examinations, and MDCT data were reconstructed with three kinds of reconstruction algorithm (two high-frequencies [A and B] and one standard [C]). In reconstruction algorithm B, the effect of low- and middle-frequency space was suppressed compared with reconstruction algorithm A. As quantitative CT parameters, kurtosis, skewness, and mean lung density (MLD) were acquired from a frequency histogram of the whole lung parenchyma in each reconstruction algorithm. To determine the difference of quantitative CT parameters affected by reconstruction algorithms, these parameters were compared statistically. To determine the relationships with the disease severity, these parameters were correlated with PFTs. In the results, all the histogram parameters values had significant differences each other (p < 0.0001) and those of reconstruction algorithm C were the highest. All MLDs had fair or moderate correlation with all parameters of PFT (-0.64 < r < -0.45, p < 0.05). Though kurtosis and skewness in high-frequency reconstruction algorithm A had significant correlations with all parameters of PFT (-0.61 < r < -0.45, p < 0.05), there were significant correlations only with diffusing capacity of carbon monoxide (DLco) and total lung capacity (TLC) in reconstruction algorithm C and with forced expiratory volume in 1 s (FEV1), DLco and TLC in reconstruction algorithm B. In conclusion, reconstruction algorithm has influence to quantitative assessments on chest thin-section MDCT examination in interstitial pneumonia patients.
UR - http://www.scopus.com/inward/record.url?scp=77950037020&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=77950037020&partnerID=8YFLogxK
U2 - 10.1016/j.ejrad.2009.02.006
DO - 10.1016/j.ejrad.2009.02.006
M3 - Article
C2 - 19324507
AN - SCOPUS:77950037020
SN - 0720-048X
VL - 74
SP - 142
EP - 146
JO - European journal of radiology
JF - European journal of radiology
IS - 1
ER -