メインナビゲーションにスキップ 検索にスキップ メインコンテンツにスキップ

Reference values of serum and urine creatinine, and of creatinine clearance by a new enzymatic method

  • O. Sugita
  • , K. Uchiyama
  • , T. Yamada
  • , T. Sato
  • , M. Okada
  • , K. Takeuchi

研究成果: ジャーナルへの寄稿 › 学術論文 › 査読

抄録

A new, totally enzymatic procedure for the determination of creatinine in serum and urine, using creatinine amidohydrolase, creatine amidinohydrolase, sarcosine oxidase and formaldehyde dehydrogenase is described. The assay was adapted to a discontinuous analyser with each analysis requiring only 20 μL of serum or 3 μL of urine. Analytical recovery of creatinine in serum and urine averaged 100·6%. Within-run and between-run precision studies gave coefficients of variation of 1·1% and 1·8%, respectively, for a serum with mean values of 83 μmol/L (9·4 mg/L) creatinine. Creatinine concentrations in serum and urine were measured by this procedure, in Japanese children and adults. The reference intervals for serum creatinine concentrations in adults were 55–96 μmol/L (6·2–10·9 mg/L) in men and 40–66 μmol/L (4·5–7·5 mg/L) in women, and for urine, 9·46–19·01 mmol/day (1070–2150 mg/day) in men and 6·75–10·61 mmol/day (764–1200 mg/day) in women. The reference intervals of creatinine clearance were 88·0–176·4 mL/min in men and 75·7–173·0 mL/min in women.

本文言語英語
ページ(範囲)523-528
ページ数6
ジャーナルAnnals of Clinical Biochemistry
巻29
号5
DOI
出版ステータス出版済み - 09-1992
外部発表はい

All Science Journal Classification (ASJC) codes

  • 臨床生化学

フィンガープリント

「Reference values of serum and urine creatinine, and of creatinine clearance by a new enzymatic method」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。