メインナビゲーションにスキップ 検索にスキップ メインコンテンツにスキップ

Study of feasible and safe condition for total body irradiation using cardiac implantable electronic devices

  • Hiroaki Matsubara
  • , Takatomo Ezura
  • , Yaichiro Hashimoto
  • , Kumiko Karasawa
  • , Teiji Nishio
  • , Masato Tsuneda

研究成果: ジャーナルへの寄稿 › 学術論文 › 査読

6   !!Link opens in a new tab 被引用数 (Scopus)

抄録

Cardiac implantable electronic devices (CIEDs) were believed to have a tolerance dose and that direct irradiation has to be avoided. Thus, no clinical guidelines have mentioned the feasibility of total body irradiation (TBI) with a CIED directly. The purpose of this work was to study a feasible and safe condition for TBI using a CIED. Eighteen CIEDs were directly irradiated by a 6-MV X-ray beam, where a non-neutron producible beam was employed for the removal of any neutron contribution to CIED malfunction. Irradiation up to 10 Gy in accumulated dose was conducted with a 100-cGy/min dose rate, followed by up to 20 Gy at 200 cGy/min. An irradiation test of whether inappropriate ventricular shock therapy was triggered or not was also performed by using a 6-MV beam of 5, 10, 20 and 40 cGy/min to two CIEDs. No malfunction was observed during irradiation up to 20 Gy at 100 and 200 cGy/min without activation of shock therapy. These results were compared with typical TBI, suggesting that a CIED in TBI will not encounter malfunction because the prescribed dose and the dose rate required for TBI are much safer than those used in this experiment. Several inappropriate shock therapies were, however, observed even at 10 cGy/min if activated. The present result suggested that TBI was feasible and safe if a non-neutron producible beam was employed at low dose-rate without activation of shock therapy, where it was not inconsistent with clinical and non-clinical data in the literature. The feasibility of TBI while using a CIED was discussed for the first time.

本文言語英語
ページ(範囲)1006-1014
ページ数9
ジャーナルJournal of Radiation Research
巻62
号6
DOI
出版ステータス出版済み - 11-2021
外部発表はい

All Science Journal Classification (ASJC) codes

  • 放射線
  • 放射線学、核医学およびイメージング
  • 健康、毒物学および変異誘発

フィンガープリント

「Study of feasible and safe condition for total body irradiation using cardiac implantable electronic devices」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル