抄録
In magnetic resonance imaging (MRI), we can use sensitivity encoding (SENSE) as a parallel imaging reconstruct from aliasing based on the sensitivity map of each coil element by reference scan. We researched SENSE reconstruction depending on the position of the reference scan. 1.5 T MRI (Achiva, Philips), 32-channel SENSE cardiac coil and rectangular solid phantom (nickel chloride) were used. Seven slices (A to G: from head to foot) were set at the phantom, slice thickness was 1 cm, and each slice interval was 5 cm. When T2 weighted axial images (TR: 3000 ms, TE: 120 ms) were taken, the center of the reference scan (C-Ref) was changed A to G, SENSE factor was changed 1 to 5, and number of signals averaged (NSA) was changed 1 to 3 with fold over suppression. Signal-to-noise ratio (SNR) and coefficient of variation (CV) from 9 regions of interest (ROI) of axial images were calculated. The SNR of slices A and G were lower than the other slices. %CV of slices F with C-Ref D was 69.4% lower than C-Ref A. %CV of slices D to F were higher than the other slices when C-Ref was A. %CV had no relation with SNR. Therefore, a 30 cm area around the center of reference can obtain best SENSE reconstruction.