The cause of artefactual ST segment distortions occurring in ambulatory ECG monitoring systems

A. Yokoi, S. Suzuki, K. Iwao, M. Okajima

研究成果: Article

抜粋

The frequency characteristics of the ambulatory ECG monitoring systems equipped with direct-recorded magnetic tape recorder were found much worse than that of ordinary ECG equipment both in gain and phase responses. In order to investigate which of these two responses plays major role in generating false ST deviation, a computer program was run with several frequency response models. The actual ECG waveforms were fed into the fictitious frequency responses (mathmatical filters) through a digitizer. The artefactual distortions of ST segment in the reproduced ECG were clarified to be caused by the phase responses of the systems with direct-recorded magnetic tape recorder

元の言語English
ページ(範囲)226-231
ページ数6
ジャーナルJapanese Journal of Medical Instrumentation
60
発行部数5
出版物ステータスPublished - 01-01-1990

All Science Journal Classification (ASJC) codes

  • Medicine (miscellaneous)

フィンガープリント The cause of artefactual ST segment distortions occurring in ambulatory ECG monitoring systems' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用