メインナビゲーションにスキップ 検索にスキップ メインコンテンツにスキップ

Toxoid flocculation assay by laser light-scattering

  • Masaaki Iwaki
  • , Yoshinobu Horiuchi
  • , Takako Komiya
  • , Tadashi Fukuda
  • , Yoshichika Arakawa
  • , Motohide Takahashi

研究成果: ジャーナルへの寄稿 › 学術論文 › 査読

6   !!Link opens in a new tab 被引用数 (Scopus)

抄録

Internationally accepted designations of antigen content for toxoid vaccines are provided by the WHO in Lf (limes flocculationis) units, based on the formation of antigen-antibody complexes. The current assay method for Lf determination involves observation of the complexes by eye, making the development of a more objective system highly desirable. Here we report a novel detection system using a laser light-scattering platelet aggregometer. The system was highly reproducible and more objective than the current method. Only three sets of duplicate data were sufficient for statistically significant determination of toxoid Lf by parabolic regression.