Twenty-year outcome of aortic valve replacement with st. Jude medical mechanical valves in Japanese patients

  • Kenji Minakata
  • , Shiro Tanaka
  • , Yohei Okawa
  • , Tatsuo Kaneko
  • , Shuichi Okonogi
  • , Akihiko Usui
  • , Tomonobu Abe
  • , Nobushige Tamura
  • , Shigeki Yanagi
  • , Ryuzo Sakata

研究成果: ジャーナルへの寄稿 › 学術論文 › 査読

17   !!Link opens in a new tab 被引用数 (Scopus)

抄録

Background: Bioprostheses have become increasingly popular for aortic valve replacement (AVR) in recent years, but mechanical valves are still the standard choice, especially for younger patients. The aim of this study was to assess the very long-term outcomes in Japanese patients who underwent AVR with St. Jude Medical (SJM) mechanical valves. Methods and Results: From 1991 to 2001, a total of 816 patients underwent AVR with SJM mechanical valves in 5 hospitals. Of these, 801 patients (mean age, 58.3±11.7 years) were analyzed in this study. There were 24 inhospital deaths (3.0%). Mean follow-up duration was 11.6±6.7 years and the 10-year follow-up rate was 84.1%. Freedom from valve-related death at 5, 10, 15, and 20 years was 96.2%, 92.7%, 88.8%, and 86.6%, respectively. The linearized ratio of major bleeding events and thromboembolic events was 1.1% per patient-year and 1.0% per patient-year, respectively. Freedom from reoperation for the aortic prosthesis was 98.0% and 94.8% at 10 and 20 years, respectively. Conclusions: The SJM mechanical valve provided excellent long-term freedom from valve-related death and reoperation in patients undergoing AVR. Therefore, this valve should be recommended to younger patients who wish to avoid reoperation.

本文言語英語
ページ(範囲)2380-2388
ページ数9
ジャーナルCirculation Journal
巻79
号11
DOI
出版ステータス出版済み - 23-10-2015
外部発表はい

All Science Journal Classification (ASJC) codes

  • 循環器および心血管医学

フィンガープリント

「Twenty-year outcome of aortic valve replacement with st. Jude medical mechanical valves in Japanese patients」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル